MCCT LABORATORY
Mechanical and Climatic Calibration and Testing Laboratory
SERVICES - CALIBRATION
C. PRESSURES
Calibration of pressure gauges and pressure transducers:
Within the LEIMC, the calibration of pressure gauges and pressure transducers is carried out in the range: 0÷700 bar.
manometer
pressure transducer
Normative references:
​
-
PV 233-05 "Metrological verification of pressure transducers";
-
NTM 3-130:94 Metrological verification of standard manometers and vacuum gauges with elastic element;
-
PSpE-LEIMC 07 "Calibration of pressure transducers";
-
PSpE-LEIMC 10 "Calibration of manometers".
Working standards:
​
-
Standard digital manometer, type Additel680, series 21816210004, range: 0÷20bar, accuracy class: 0.05;
-
Standard digital pressure gauge, type Additel680, series 21816220040, range: 0÷700bar, accuracy class: 0.05.
Measuring equipment:
​Pressure pump type ADDITEL.927, Series 13215050012